WLN-CA.W.33.4Determination of the total phosphates content (low level) in water by means of the auto analyser.in-house method
Principle:Ammonium molybdate in an acid environment with orthophosphate ions, antimony oxyions and ascorbic acid will form a blue coloured compound. The intensity of the blue colour is measured at 880 nm and is a measure for the phosphate content in question.
Determination:@Ortho-fosfaat, laag niveau
Performance characteristics:
Reporting limit:1 ug P/l, matrix dependent
Measurement uncertainty:25%
Sustainability:28 days
Sampling bottle:Plastic , white , 250 ml